BS EN 13823 SBI Building Materials Single Burning Item Combustion Equipment
The Single Burning Item Test Equipment for Building Materials is a testing device independently developed and manufactured by our company in accordance with EN 13823. This apparatus determines the fire reactivity of building materials or articles (excluding flooring materials) in a single-body burning test (SBI).
The system employs a computer interface for direct control of testing equipment, significantly saving manpower and resources. The test program features an intuitive interface written in Visual Basic 6.0, with data output in Excel format for easy processing. Test data curves can be saved and printed in real time, and complete test records are securely stored.
Compliance Standards
- EN 13823 "Fire reaction of building products - Test methods for individual combustion of building products without underlayment"
- GB/T 20284-2006 "Individual combustion tests of building materials or products"
- GB 8624-2012 "Classification of flammability of building materials and products"
Technical Specifications
| Parameter |
Specification |
| Name |
Single Burning Items |
| Standard |
BS EN 13823 |
| Combustion Chamber Dimensions |
3000mm × 3000mm × 2500mm |
| Exhaust Duct Dimensions |
5200mm × 3000mm × 1800mm |
| Control Box Dimensions |
78mm × 55mm × 1800mm |
| Exhaust Velocity |
0.50m³/s to 0.65m³/s |
| Temperature Measurement |
K-type thermocouple (±0.5°C accuracy) |
| Oxygen Analysis |
0-25% range, <12s response |
| CO₂ Measurement |
0-10% range, >1% linearity |
| Power Supply |
AC 220V 50Hz, 15kW max |
Key Features
- High-precision pressure sensor for flue gas duct with ±2Pa accuracy
- Test duration: 1560 seconds
- Oxygen analysis (paramagnetic): 0-25% range with <12s response time
- Carbon dioxide measurement (IR type): 0-10% range with >1% linearity
- Data acquisition output resolution: ≥100x10
- Colorimetric standard accuracy: ±5%
- Output linearity (transmittance): <3%
- Control cabinet power supply: AC220V 50HZ
- Equipment power: 15kW
- Safety alarm activates when flue gas duct temperature exceeds 400℃