X-RAY SOURCE 130 kV MICROFOCUS X-RAY SOURCE L9181-02
FEATURES
●Focal spot size:5μm(at 4W)
The focal spot of 5μm of the sealed type X-ray tube offers sharpand clearX-ray images even ata high magnifcation.
●No high voltage cable connection required
High voltage power supply is integrated.
●External control via RS-232C interface
OHigh power:Maximum output 39 W
APPLICATIONS
●Non-destructive inspection
●X-ray CT
[Applicable objects]
·Electronic component
·Printed circuit board
·Plastic component
·Metal component
| SPECIFICATIONS | ||||
| GENERAL | ||||
| Parameter | Description/Value | Unit | ||
| X-ray tube voltage setting range | 0 to 130 | kV | ||
| X-ray tube current setting range | 0 to 300 | μA | ||
| X-ray tube voltage operational range ① | 40 to 130 | kV | ||
| X-ray tube current operational range ① | 10 to 300 | μA | ||
| Maximum output | Small focus mode | 8 | W | |
| Middle focus mode | 16 | |||
| Large focus mode | 39 | |||
| X-ray focal spot size (Nominal value) | Small focus mode | 8(5μm at 4 W) | μm | |
| Middle focus mode | 20 | |||
| Large focus mode | 40 | |||
| X-ray output window material/Thickness | Beryllium/0.2 | mm | ||
| X-ray beam angle② | Approx.45 | degree | ||
| Focus to object distance(FOD) | Approx.13 | mm | ||
| Terget material | Tungsten | 一 | ||
| Weight③ | Approx.10.5 | kg | ||
| Communication method | Interface:RS-232C(9-pin D-sub connector) | 一 | ||
| RATINGS | ||||
| Parameter | Description/Value | Unit | ||
| Input voltage(DC) | +24(+2.4,-0) | V | ||
| Power consumption | Less than 120 | W | ||
| Read output | Continuous rating | — | ||
| Operating ambient temperature | +10 to +40 | ℃ | ||
| Storage ambient temperature | 0 to +50 | ℃ | ||
| Operating and storage humidity | 20 to 85(No condensation) | % | ||
| REGULATION AND STANDARD | ||||
| Parameter | Description | Unit | ||
| RoHS Directive | EN 50581 Category 9 | 一 | ||
| EMC | IEC/EN 61326-1 | Emission limits:CISPR 11 Group 1 Class A Immunity requirements:Table 2 | 一 | |
| CONTROL SOFTWARE | ||||
| Parameter | Description | Unit | ||
| Applicable PC | PC/AT compatible | 一 | ||
| Applicable OS | Windows®XP,7 | 一 | ||
| RS-232C | 一 | |||
| NOTE:①See the graph of the X-ray tube voltage /current operation range. | ||||
| ②Reference value:With 50%of maximum X-ray emission. ③This weight includes the accessories of approx.0.25 kg. | ||||
| ④The control software is provided as a sample software for the purpose of MFX operation. The performance of the software is not guaranteed. | ||||