| Function Module |
Description |
| Operation |
Keyboard and mouse can finish all operations |
| X-ray tube control |
Using mouse to click the X button can turn on or off the X-ray. The real-time tube voltage and current value will display beside, users can click up and down button, or drag the slider, or type to adjust. |
| Status bar |
Indicates the interlock status, pre-heat status, and X-ray status by flashing alternately red and green. |
| Image effect adjustment |
The brightness, contrast and gain of the image can be adjusted freely to achieve a satisfactory result. |
| Product list |
Users can save the inspection parameters such as Z-axis position, brightness, contrast and gain, and can directly call the parameters when inspecting the same product, to improve the inspection efficiency. |
| Navigation window |
After the camera takes a photo of the table, click anywhere in the photo, the table will move to the place you click and display on the screen. |
| Motion axis status |
Display real-time coordinates. |
| Inspection result |
The measurement results (voids rate, distance, area, etc., set by users) display in order. |
| Speed control |
The movement speed of each axis can be adjusted to slow, normal and fast. |
| Voids rate measurement |
Automatic calculation, parameters adjustment, add voids manually, and save parameters functionality for precise void analysis. |
| Other measurement function |
Distance, distance rate, angle, radius, and perimeter measurements for comprehensive component analysis. |
| Automatic inspection |
Manual setting, array, and automatic identification modes for efficient inspection workflows. |