System Introduction of IC6200 Ion Chromatography Machine
Built for demanding analytical environments, the IC6200 ion chromatography machine incorporates high- and low-pressure protection mechanisms, leak detection with automatic shutdown, and an auto-range conductivity detector based on DSP architecture. These safety and reliability features protect the analytical system and ensure uninterrupted operation in regulated laboratories.
System Parameters of IC6200 Ion Chromatography Machine
|
System Parameter
|
|
|
Qualititative Repeatability
|
≤0.05%
|
|
Quantitative Repeatability
|
≤0.2%
|
|
Pump System
|
|
|
Flow Rate (Full PEEK Pump)
|
(0.001–12.000) mL/min
|
|
Pump Head Pressure Resistance
|
(0–42) MPa
|
|
Flow Rate Set Error
|
≤ ±0.1%
|
|
Flow Rate Stability
|
≤ ±0.1%
|
|
Pressure Display Resolution
|
0.01 MPa
|
|
Suppressor
|
|
|
Suppressor Type
|
Electrolytic self-regenerating membrane suppressor
|
|
Constant Current Range
|
0–500 mA, adjustable in increments of 1 mA
|
|
Column Oven
|
|
|
Temperature Range
|
Ambient +5°C to 85°C
|
|
Temperature Stability
|
±0.01°C
|
|
Compatibility
|
Compatible with 250 mm and 150 mm length columns
|
|
AS2800 Series Autosampler
|
|
|
Quantitative Repeatability
|
< 0.5% RSD (full-loop injection)
|
|
Sample Positions
|
48 positions / 11 positions
|
|
Wash Ports
|
2 ports (fine wash and coarse wash)
|
|
Operating Temperature
|
10–40°C
|
|
Vial Types
|
5mL/ 100mL
|
|
Injection Volume
|
10 µL – 4000 µL (depending on sample loop size)
|
|
Single Injection Cycle Time
|
Approx. 20s (depending on wash time and injection volume)
|
|
Injection Mode
|
Full-loop injection
|
|
AS3110 Series Autosampler
|
|
|
Sample Capacity
|
64/120 positions
|
|
Cooling Temperature Range
|
4–30°C
|
|
Repeatability
|
< 0.3% RSD (full-loop injection, injection volume > 5 µL)
|
|
Syringe Specifications
|
Standard: 500 µL; Optional: 250 µL, 1000 µL, 2500 µL
|
|
Sample Loop Specifications
|
Standard: 100 µL; Optional: 25 µL, 50 µL, 200 µL
|
|
Carryover (Cross Contamination)
|
< 0.01%
|