HbA1c Analyzer LD-100 POCT HbA1c Analyzer Boronate Affinity Portable Analyzer For HbA1c
Intended Use
LLD-100 POCT Analyzer is intended for the percent determination of hemoglobin A1c (HbA1c) in human whole blood which is a semi-automatic instrument used to analyze glycated hemoglobin. LD-100 is small and light, it can be used to detect of HbA1c in medical units at all levels. Fually Automatic system can help the test half work with double results.
Feature
- Protable, easy to carry
- Fast test speed, WIthin 3 minutes
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color touch screen
- External printer (Blue-tooth)
- USB Port, LIS Support

Specification
| Model | LD-100 Analyzer |
| Method | Bornate Affinity |
| Measurement Time | ≤3 min |
| Sample Type | Whole blood |
| Shelf-Life of test kit | 9 Months |
| Weight | 800gs |
Reliable
| IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
| Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
| CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Operation
1. Place cartridge in analyzer.
2. Push bead into reagents.
3. Insert sample into cartridge.
Performance Parameters
- Accuracy: Is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
- Functions of the analyzer:The function of report output in English; Abnormal alarm function.