HbA1c Test Analyzer Hemoglobin Detection Analysis LD-100 Poct HbA1c Fully Automatic Testing Analyzer
Intended Use
The instrument is LD-100(POCT) One-Step HbA1c Analyzer, it's intended for the percent determination of hemoglobin A1c (HbA1c) in human
whole blood. LD-100 HbAc Analyzer is composed of the systems of sample cup mixing, colorimetric detection and system control, result display which can fast testing Hemoglobin with less time. Because the LD-100 HbA1c Analyzer is small and light so it 's very protable and suitable for the detection of HbA1c in medical units at all levels.
Specifications
Model | LD-100 |
Sample Type | Whole blood |
Sample Volume | 4ul |
Testing Time | Within 3 minutes |
Data storage | 7000 |
Power connection | Detachable power supply cord |
System Compose | Sample cup mixing, Colorimetric detection And System control, Result display |
Device category: | Fixed |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Voltage | 12VDC-3.33A |
Analytical wave-length | 510nm and 580nm. |
Certification | NGSP, IFCC |

PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the buffer and mixed to release the hemoglobin from the red blood cells.
- The total concentration of hemoglobin depends on the initial reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and glycosylated hemoglobin is determined by detecting the decrease of fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can be determined.
Feature
- One-Step Testing,Fully Automatic System
- Protable and convenient, Testing Anytime , Anywhere
- Fast test speed and High Accuracy within 3 min
- Battery Integrated
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color touch screen just need to clik the screen and easy to read
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge

Performance Parameters
- Accuracy: Is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
- Functions of the analyzer:The function of report output in English; Abnormal alarm function.