The ultrasonic hardness tester is using ultrasonic contact impedance method to compare and test patterns, this is an innovative design of auto-calibration of probe, presents advantages of high precision, fast speed, easy operation, etc.
3 User materials with setup guide was included, in addition, the hardness measurement covering most ferrous metals, non-ferrous metals and their alloys in industrial production.
It is applicable in hardness test of flange edges, root of gear groove, stamping parts, thin plates, for surface-hardened gear grooves and its taper parts, as well as shafts, thin-walled pipes, turbines.etc.
Features
- Accord to the Testing Standard GB/T 34205-2017
- For materials with different elastic modulus, provide multi-point calibration function for any hardness standard based on your standard hardness samples.
- Support hot-swap and self-calibration: automatically + by manual identifies the probe
- The material function includes 3 new material settings to facilitate hardness testing of special materials
- The display adopts IPS (In-Plane Switching) full digital LCD display with wide viewing angle, and the back light can be adjusted arbitrarily
- USB communication interface allows for data exchange with PC and to set up on PC.
- The upper and lower limits of the hardness value can be pre-set, and an automatic alarm will be issued if the range is exceeded, which is convenient for users to meet the needs of batch testing
- 600 sets of hardness measurement data each includes measurement date, sensor type, material, number of times, single measurement value, average value, maximum value, minimum value, hardness scale and other information
Technical parameters
| Measurement direction |
perpendicular to testing surface |
| Hardness scale |
Vickers (HV), Brinell (HB), Rockwell (HRB), Rockwell (HRC), Rockwell (HRA), |
| Display Measurement |
Vickers (HV) + selected hardness scale |
| Measuring materials |
steel and cast steel, cast aluminum alloy, pure copper. |
| Display |
3.5-inch IPS full digital color LCD display, resolution 480*320. |
| Probe recognition |
automatic recognition + manual setting, supports hot swapping |
| Data storage |
maximum 600 measurement |
| Working voltage |
3.7V lithium-ion battery pack |
| Continuous working time |
about 10 hours |
| Charging power supply |
5V/1000mA; |
| Communication interface |
USB2.0 |
| Probe test force |
2kgf (optional, 5kgf, 10kgf) |
| Test force |
Application |
| 98N |
Forgings, welds and heat affected zones |
| 50N |
Hardened quenching layer or carburized layer of machining parts: camshafts, turbines, welds and heat-affected zones |
| 20N |
Flanges, gear groove, stamping parts, molds, crankshafts, turbine rotors, etc. |
| 10N |
Ion nitriding layer of stamping dies, forgings |
| 8N |
Precision parts, reducers and bearing rotating rings |
| 3N |
Hardness test of copper layer or chromium coating (plating) (t≥0.04mm) on cylinder , and coating or hardened quenching layer (t≥0.02mm) on low pressure cylinder |
| 1N |
Thin layer, polished surface |
Standard Package
Main unit, Probe (2kgf), Probe cable, 5V ac power adapter, Standard test block, Carrying case
Table 1
| Maximum allowable relative error |
Hardness of standard test block |
Maximum allowable relative error of hardness tester +/-% |
| HV1 |
HV2 |
HV5 |
HV10 |
HV0.1 |
HV0.3 |
HV0.8 |
|
<250HV |
4 |
4 |
4 |
4 |
5 |
5 |
4 |
| 250HV-500HV |
4 |
4 |
4 |
4 |
6 |
6 |
4 |
| >500HV-800HV |
5 |
4 |
4 |
5 |
7 |
7 |
5 |
| >800HV |
6 |
4 |
4 |
4 |
8 |
8 |
6 |
Table 2
| Maximum allowable relative repeatability |
Hardness of standard test block |
Maximum allowable relative repeatability of hardness tester % |
| HV1 |
HV2 |
HV5 |
HV10 |
HV0.1 |
HV0.3 |
HV0.8 |
|
≤250HV |
8 |
5 |
5 |
5 |
8 |
8 |
9 |
| >250HV |
6 |
5 |
5 |
5 |
6 |
6 |
6 |
Table 3_Probe specification
| Probe |
1kgf manual probe |
2kgf manual probe |
5kgf manual probe |
| Accessory Type |
Optional accessory |
Standard accessory |
Optional accessory |
| Test force |
10N |
20N |
50N |
| Diameter |
22mm |
22mm |
22mm |
| Length |
150mm |
150mm |
150mm |
| Resonant rod diameter |
2.4mm |
2.4mm |
2.4mm |
| Maximum roughness of test surface |
Ra<3.2um |
Ra<5um |
Ra<10um |
| Minimum workpiece weight |
0.3kg |
0.3kg |
0.3kg |
| Minimum workpiece thickness |
2mm |
2mm |