| Key feature and specification comparison | 8588A | 8558A | 
| DC voltage | 100 mV – 1000 V | 100 mV – 1000 V | 
| AC voltage | 10 mV – 1000 V, 1 Hz – 10 MHz | 10 mV – 1000 V, 1 Hz – 10 MHz | 
| Resistance, LoI, HV | 1 Ω - 10 GΩ | 1 Ω - 10 GΩ | 
| DC current | 10 uA – 30 A | 10 uA – 2 A | 
| AC current | 10 uA – 30 A, 1 Hz to 100 kHz | 10 uA – 2A, 1 Hz to 100 kHz | 
| Digitize V | 100 mV - 1000 V, 5 MSamples/s, up to 20 MHz BW | 100 mV - 1000 V, 5 MSamples/s, up to 20 MHz BW | 
| Digitize I | 10 uA - 30 A, 5 MSamples/s, up to 4 MHz BW | 10 uA - 2 A, 5 MSamples/s, up to 4 MHz BW  |  		
| Frequency (V, I, BNC) | 1 Hz to 10 MHz, 1 Hz to 100 kHz, 10 Hz to 100 MHz | 1 Hz to 10 MHz, 1 Hz to 100 kHz, 10 Hz to 100 MHz | 
| Temperature | PRT / Thermocouple (ext CJC) | PRT / Thermocouple (ext CJC) | 
| Capacitance | 1 nF – 100 mF | No | 
| RF power | R&S NRP Series | No | 
| Ext. dc current and ac current | A40B current shunt and other shunts | No | 
| Graphical display | Yes | Yes | 
| Visual Connection Management ® | Yes | Yes | 
| Programmable front/rear input switching | Yes | Yes | 
| Ratio difference, deviation - ohms, voltage, current | Yes | Yes | 
| External 10 MHz ref clock, 50 Ω/Hi-Z | Yes | Yes | 
| A40B and other shunt asset management | Yes | No | 
| GPIB 488.2, Ethernet, USB TMC | Yes | Yes | 
| SCPI command compatibility | Yes | Yes | 
| 3458A emulation, 8508A emulation | Yes | Yes | 
| Volatile memory | 15 million | 15 million | 
| Level and other trigger | Yes | Yes | 
| FFT onboard | Yes | No | 
| Reading rates: 5 MS/s into memory, bus: 100 kS/s at 4.5 digits, 1 rdg/s at 8.5 digits | Yes | Yes | 
| Function | 8588A uX/X of reading + uX/X of range | 8558A uX/X of reading + uX/X of range | ||||
| 95% | 99% | 95% | 99% | |||
| DC voltage | 10 V | relative | 2.7 + 0.05 | 3.5 + 0.06 | 4.0 + 0.06 | 5.2 + 0.08 | 
| absolute | 2.8 + 0.05 | 3.6 + 0.06 | 4.1 + 0.06 | 5.3 + 0.08 | ||
| AC voltage | 10 V, 1 kHz | relative | 60 + 5 | 77 + 6.5 | 80 + 10 | 103 + 13 | 
| absolute | 64 + 5 | 83 + 6.5 | 90 + 10 | 116 + 13 | ||
| Resistance | 10 kΩ | relative | 7 + 0.5 | 9 + 0.6 | 10 + 0.6 | 13 + 0.7 | 
| absolute | 7.2 + 0.5 | 9.2 + 0.6 | 10.3 + 0.6 | 13.3 + 0.7 | ||
| DC current | 1 mA | relative | 6.5 + 4 | 8.4 + 5 | 9 + 5 | 12 + 6 | 
| absolute | 7.6 + 4 | 10 + 5 | 9.8 + 5 | 13 + 6 | ||
| AC current | 1 mA, 1 kHz | relative | 250 + 50 | 323 + 65 | 300 + 100 | 387 + 129 | 
| absolute | 260 + 50 | 335 + 65 | 310 + 100 | 400 + 129 | ||
| Frequency | BNC,1 kHz | relative | 0.5 uHz/Hz | 0.5 uHz/Hz | 0.5 uHz/Hz | 0.5 uHz/Hz | 
| Temperature | PRT 100 Ω | relative | ± 5 mK | ± 5 mK | ± 5 mK | ± 5 mK | 
| Capacitance | 1 uF | relative | 400 + 100 | 516 + 129 | N/A | N/A | 
| absolute | 406 + 100 | 523 + 129 | N/A | N/A | ||
 95% 1 year relative specification. Fluke Calibration guarantees to 99% confidence interval k=2.58. For 99%, multiply by (2.58/2).
Bench Multimeter Selection Guide
| Model | Resolution | Accuracy | Application | 
| 8588A | 8.5 | 2.7 µV/V 95% confidence level, 1 year relative accuracy | Metrology | 
| 8558A | 8.5 | 4.0 µV/V 95% confidence level, 1 year relative accuracy | High-Speed Automation | 
| 8845A/8846A | 6.5 | Basic V dc accuracy of up to 0.0024 % | Bench | 
| 8808A | 5.5 | Basic V dc accuracy of 0.015 % | Bench |