Techncial Data of LC3400 High Performance Liquid Chromatography Instrument
| Product Name | LC3400 Liquid Chromatography Instrument |
| Maximum Operating Pressure | 62MPa/75MPa |
| Flow Rate Setting Range | 0.0000~12.0000mL/min |
| Flow Rate Setting Increment | 0.0001mL/min |
| Gradient Error | ≤ 0.75% |
| Gradient Stability | 0.2% RSD (5% to 95%, 0.2 to 2.0mL/min) |
| Flow Rate Setting Value Error | ±0.20% |
| Flow Rate Stability | ≤0.075% RSD |
| Autosampler | |
| Injection Volume | 100μL, 200μL, 500μL, 1000μL, 1500μL, 2000μL |
| Repeatability | <0.3%RSD (2μL~Max), <1.0%RSD (0.1μL~1.9μL); |
| Linear Correlation Coefficient | >0.9999 |
| Number of Sample Position | 108 (2*54, 2mL), 96-well plates |
| Temperature Setting Range | 4.0~40.0℃ (optional refrigeration) |
| Temperature Setting Value Error | ±0.5℃ (optional refrigeration); |
| Temperature Stability | ±1℃ (optional refrigeration) |
| Column Oven | |
| Temperature Setting Range | 4℃-90℃ |
| Temperature Stability | ±0.03℃/h |
| Temperature Setting Value Error | ±0.1℃ |
| UV Detector | |
| Baseline Noise | ≤8.0×10-6 Au (1mL/min, methanol, C18) |
| Baseline Drift | ≤1.0×10-4 Au/h (1mL/min, methanol, C18) |
| Minimum Test Concentration | ≤5×10-9 g/mL (naphthalene/methanol) |
| Spectral Range | 188-900nm |
| Wavelength Error | ≤ ±1nm |
| Wavelength Repeatability | ≤ ±0.1nm |
| Linear Range | ≥1×104 or 2.5Au |
Performance of High Performance Liquid Chromatography Instrument
1. High-precision ultra-high-pressure infusion unit
2. Low system diffusion volume and optimized flow path
3. Needle-in-flow autosampler with ultra-low injection cycle time
4. Extremely low cross-contamination
5. Active preheat function, fast warming rate