| Parameter | Value |
| Average heating rate | From 50℃ -90℃, ≥ 3.5℃ / s |
| Maximum heating rate | From 50℃ -90℃, ≥ 5.0℃ / s |
| Average cooling rate | From 90℃ -50℃, ≥ 3.5℃ / s |
| Maximum cooling rate | From 90℃ -50℃, ≥ 5.0℃ / s |
| Module temperature control accuracy | ≤0.4℃ |
| Temperature accuracy | ≤0.4℃ |
| Temperature uniformity | ±0.8℃ |
| Repetitiveness | The CV of the Ct value should not be greater than 3% |
| Sample linearity | Linear regression coefficient was r≥ 0.980 |
| Fluorescence linear | Linear regression coefficient was r≥ 0.990 |