| Display range of volume delivered | | | 0.1~9999.9mL (in steps of 0.1 mL,1000 mL/h or above, in steps of 1mL) | | | | | | | ||||||||
| Setting of delivery limit | | | 0.1~9999.9mL (in steps of 0.1 mL, 1000 mL/h or above, in steps of 1mL) | | | | | | | ||||||||
| Accuracy of flow rate | | | Accuracy including syringe: within ±2% (Mechanical accuracy: within ±1%) | | | | | | | ||||||||
| KVO-Keep vein open rate | | | 0.1~5mL/h adjustable, 1ml/h as default; | | | | | | | ||||||||
| Occlusion pressure | | | 0.1~1.2bar, 9 adjustable levels; | | | | | | | ||||||||
| Alarms | | | End of infusion --infusion ends soon / Occlusion/ Plunger or clutch disengaged/ Syringe dislocated/ Low battery--end of battery/ Power cable disconnected/ Non--operation time over/ Flow rate over limit/ Volume complete (a volume limit has been set)/ System error/ Deactivation of KOR rates possible | | | | | | | ||||||||
| Power supply | | | AC 110~230V, 50/60Hz; Inner Lithium battery: DC 11.1V; Continuous operation time of battery: over 6 hours for dual channel when battery is recharged for 12 hours under power off condition | | | | | | | ||||||||
| Power consumption | | | 20VA | | | | | | | ||||||||
| Operation conditions | | | Ambient temperature: 5~40 ℃ ; Atmospheric pressure: 86~106kPa; Relative humidity: 20%~80%; Non-‐condensable; | | | | | | | ||||||||
| Storage and transportation conditions | | | Ambient temperature: --20~50℃; Relative humidity: 10%~95%; Non--condensable; Atmospheric pressure: 50~106kPa; | | | | | | | ||||||||
| Classification | | | Class II, Type CF | | | | | | | ||||||||
| IP-Protection | | | IP22 | | | | | | | ||||||||
| Dimension | | | 310×120×130mm (W×H×D) | | | | | | | ||||||||
| Maximum weight | | | Less than 2.5kg | | | | | | | ||||||||
| Operation type | | | Continuous operation | | | | | | | ||||||||