This product is a battery-powered, auto-ranging, true RMS digital multimeter with a 4000 counts LCD display.
Introduction
This product is a battery-powered, auto-ranging, true RMS digital multimeter with a 4000 counts LCD display.
Safety Information
To avoid possible electrical shock, fire, or personal injury, please read all safety information before you use the product. Please use the product only as specified, or the protection supplied by the product can be compromised.
- Examine the case before you use the product. Look for cracks or missing plastic. Carefully look at the insulation around the terminals.
- The measurement must be made within the allowable measuring range.
- Do not use the product around explosive gas, vapor, or in damp or wet environments.
- When the voltage to be measured exceeds 36V DC or 25V AC, the operator shall be careful enough to avoid electric shock.
- Misuse of mode or range can lead to hazards, be cautious. “OL” will be shown on the display when the input is out of range.
- Low level of a battery will result in incorrect readings. Change the batteries when battery level is low. Do not make measurements when the battery door is not properly placed.
1. General Specifications
| General Specifications |
| Display | 4000 counts | Ture RMS | √ |
| Ranging | Auto/Manual | Data Hold | √ |
| Material | ABS | Backlight | √ |
| Update Rate | 3 / s | Flashlight | √ |
| Low Battery Indication | √ | Auto Power Off | √ |
2. Electrical Specifications
| Electrical Specifications |
| Function | Range | Resolutio n | Accuracy | Max |
| DC VOLTAGE (V) | 4.000V | 0.001V | ±(0.5 +3) | 600V |
| | AC VOLTAGE (V) | 4.000V | 0.001V | ±(1.0 +3) | 600V |
| 40.00V | 0.01V |
| 400.0V | 0.1V |
| 600V | 1V |
| | AC mA | 999.9mA | 0.1mA | ±(2.0 +3) | 9.999A |
| | AC A | 9.999A | 0.001A |
| | DC mA | 999.9mA | 0.1mA | ±(1.0 +4) | 9.999A |
| | DC A | 9.999A | 0.001A |
| | Resistance | 4.000kΩ | 0.001kΩ | ±(1.5 +3) | 40MΩ |
| 40.00kΩ | 0.01kΩ | ±(0.5 +3) |
| 400.0kΩ | 0.1kΩ |
| 4.000MΩ | 0.001MΩ |
| 40.00MΩ | 0.01MΩ | ±(1.5 +3) |
3. Environmental Specifications
| Environmental Specifications |
| Operating | Temperature | 0~40℃ |
| | Humidity | <75% |
| Storage | Temperature | -20~60℃ |
| | Humidity | <80% |
Test for NCV
- Keep pushing the NCV button to enter the NCV mode.
- Hold the product and move it around, the built-in beeper will beep when the inner sensor detects AC voltage nearby. The stronger the voltage is, the quicker the beeper beeps while the central LED light will twinkle.
Test for Current
- Put the red lead to terminal and put the black lead to COM terminal. The current measure will be auto matched when you put the leads in.
- Touch the probes to the correct test points of the circuit to measure the current.
- Read the measured current on the display
- When you input the current over 2 AMP, the testing time should be less than 3 seconds.
- When the probes are in the current terminal but you have no operations, this product will alarm each 4 seconds to remind that you are in the current mode.