| Specification | Details |
|---|---|
| Product Series | T3030 Series |
| Power Supply Requirements | 15 15 VDC 10% 24 - 18-30 VAC RMS, 50/60 Hz 18 to 42 VDC, polarity protected |
| Power Consumption | Typical 0.7 W at nominal voltage of 24V AC RMS |
| Range | See Ordering Information for calibrated ranges |
| Resolution | Minimum 10bit or analog equivalent |
| Accuracy (Single Wavelength, 0~2,000 ppm example) | 400-1250 ppm: 30 ppm or 3% of reading, whichever is greater* ** 1250-2000 ppm: 5% of reading + 30ppm *, ** |
| Accuracy (Dual Wavelength, 0~2,000 ppm example) | 75 ppm or 10% of reading (whichever is greater) |
| Accuracy (Dual Wavelength, 0~200,000 ppm example) | 5% of reading or 0.3 vol% CO2 (whichever is larger) at 0 - 20 vol% CO2 and 0C to +25C At 25C to 50C: additional 0.1% of reading or 0.01 vol% CO2 (whichever is larger) per degree C. |
| Temperature Dependence | 0.2% FS / C |
| Response Time (tau 63%) | < 200 sec |
| Aging (Dual Wavelength Only) | +/- 5% of reading annual drift. Less than +/- 25% of reading within product life time. |
| Recommended Recalibration Interval | Single Wavelength No calibration required Dual Wavelength - 1 year |
| Working Temperature | 0 to +50C (measures within tolerance) |
| Operation Temperature (Sensor Is Powered) | -30 to +50C |
| Storage Temperature | -40 to +70C |
| Minimum Lifetime | 10 years or 30,000 operating hours |
| RH Range | 0% to 99% RH (non-condensing) |
| Atmospheric Conditions Tested | Marine, salt mist and particles present; O2 level 0-21 vol%; 0 to 100% RH; Ammonium max 200mg/m3; SO2 max 0.5 vol%; Ethylene max 500ppm vol; Ozone |
| Output Configurations | Analog: 0 to 5 V, 0 to 10 V, 4 to 20mA Digital |
| Analog Output Error | 1% |
| Example Model (Potted) | T3031-2-2K-24-P (Dual Channel, 0-2000 PPM, 0 to 10 VDC o/p, Potted) |
| Example Model (Potted) | T3032-2-5K-24-P (Dual Channel, 0-5000 PPM, 4 to 20 mA o/p, Potted) |
| Example Model (Potted) | T3032-2-10K-24-P (Dual Channel, 0-10,000 PPM, 4 to 20 mA o/p, Potted) |
| Example Model (Potted) | T3032-2-20K-24-P (Dual Channel, 0-20,000 PPM, 4 to 20 mA o/p, Potted) |