Product Name | |
Name Model | IPX9K waterproof test chamber |
1.Specimen Description | |
This Test Equipment is Suitable for: | |
2. Volumes | |
2.1 Nominal | 1000L |
2.2 Inner Box Size | 1000 mm *1000 mm *1000 mm(W*H*D) |
2.3 External Dimensions | 1500 mm*1900 mm*1500 mm(W*H*D) |
2.4 Weight | About 300kg |
3.Performance | |
3.1 Test | Ambient temperature is +25℃, relative humidity ≤85%, no specimen in the test chamber |
3.2 Test Method | 9K: water pressure: 8000 ~ 10000kPa, water temperature 80ºC ± 5ºC protection, in 30 degree increments (four high-pressure nozzles: 0, 30, 60 and 90 degrees), in each angle for 30 seconds, a total of 120 seconds (2 minutes). |
3.3 Temperature Range | 9K:RT+5~80±5℃, |
3.4 Nozzle | 9K: φ1.8mm (shaped nozzle) |
3.5 Nozzle | 9K: 4 flat beam nozzles, installed at 30 degree intervals in a 90 degree plane, 100mm to 150mm from the center, adjustable. |
3.6 Water Pressure | 9K: 8000 to 10000kpa display (100bar) |
3.7 Flow | 14~16L/min can be set |
3.8 Turntable | 0~10r/min |
3.9 Meet the | DIN 40050-9-1993 (IEC 529-1989) High pressure/vapour jet flushing tests as defined in electrical equipment for road vehicles |
4.Structural Features | |
4.1 Insulation | Inside Wall Material: stainless steel plate SUS304, mirror |
4.2 Test Chamber | Lead Hole: diameter φ50mm 1, (located at the back of the box) |
4.3 Door | Hinged single door (Plexiglas), 1pc |
4.4 Control Panel | Temperature, flow, time control display |
4.5 Machinery Room | The mechanical room contains: |
4.6 Power | Radiator fans |
4.7 Heater | Using stainless steel armored heater |
4.8 Water Supply | Circulation, water level control, import and export water valve. (To replace the water regularly) |
5.Flow (pressure) and Time Control Systems | |
5.1 Working Method | High-pressure pump → tee → nozzle and balance pressure outlet → electromagnetic flow meter (pressure gauge, 100 bar) → divided into four ways (each controlled by solenoid valve, solenoid valve by setting time → controlled by controller) → in 30 degrees of angle progression (four high-pressure nozzles: 0, 30, 60 and 90 degrees), in each angle for 30 seconds, a total of 120 seconds (2 minutes), cycle work. |
6.Water Temperature Heating System | |
6.1 Working Method | BT(H)C balance temperature regulation mode. |