P/N 989803174921 PHiliphhilip Goldway CTG7 Fetal Monitor US Probe
The CTG7 Maternal/Fetal Monitor provides reliable maternal and fetal monitoring before and during delivery. Provides comprehensive monitoring of fetal heart rate, uterine activity and fetal movement, as well as maternal ECG/HR, NIBP and SpO2/PR. The compact design and adjustable display enable monitoring in a variety of clinical situations. The monitor is portable so it can be easily moved from one patient to another using the integrated handle or mounted on a movable stand.
Recorder
| Type
| 150mm internal thermal array recorder
|
| Print speed
| 1 cm/min, 2 cm/min, 3 cm/min
|
| Print paper Standard
| Z-fold: 5 stacks per box,150 perforated sheets per stack 150 mm x 100 mm x 150 pages
|
AC Power Parameter Specification
| Mains power input
| AC 100 V – 240 V, 50Hz/60Hz
|
| Power input
| 100VA
|
| Fuse
| T1.6AH 250V
|
Environment
| Temperature - operating
| 10⁰C – 40⁰C (50⁰F – 104⁰F)
|
| Temperature - storage and transportation
| -20⁰C – 55⁰C (-4⁰F – 131⁰F)
|
| Relative humidity - operating
| ≤80% (non-condensing)
|
| Relative humidity - storage and transportation
| ≤ 95% (non-condensing)
|
| Barometric pressure - operating
| 70 kPa – 106 kPa
|
| Barometric pressure - storage and transportation
| 50 kPa – 106 kPa
|
Fetal Heart Rate
| FHR measurement and alarm range FHR accuracy
| 50 bpm – 240 bpm 50 bpm – 210 bpm; ±2 bpm 211 bpm – 240 bpm; ±4 bpm
|
| Display sweep speed
| 1 cm/min, 2 cm/min, 3 cm/min
|
| Ultrasonic frequency
| 1 MHz
|
| Mechanical index (MI)
| < 1
|
| Thermal index (TI)
| < 1
|
| Ultrasonic strength
| Pr² < 1 MPa ᴵob³ < 20 mW/cm2 ᴵspta⁴ < 100 mW/cm2
|
| Transducer protective grade
| IPX8, watertight to a depth of 1.0 m for at least five hours
|
TOCO
| Range
| 0 – 100
|
| Nonlinear precision
| ±10%
|
| Resolution
| 1
|
| Zeroing mode
| Manual
|
| Transducer protective grade
| IPX8, watertight to a depth of 1.0 m for at least five hours
|
SpO2
| SpO2 measurement range
| 0% – 100%
|
| SpO2 pulse rate range
| 30 bpm – 300 bpm
|
| SpO2 PR accuracy
| ±1 bpm or ±2%, whichever is greater
|
| Wavelength range Max optical output power
| 500 – 1000 nm for all specified sensors 15 mW for all specified sensors
|
| Updating time
| Typically 1 second
|
| PR sound
| PR tone
|
| PR display
| Numeric
|
| Display sweep speeds
| 12.5mm/s, 25mm/s, 50mm/s
|