MOX1 MOX-1 AA829-210 Oxygen Gas Sensor for Ventilator Anesthesia Machine and Oxygen Concentration Monitorin
Features:
- Wide measurement range: It can measure oxygen partial pressure from 0 mbar to 1500 mbar, corresponding to an oxygen concentration range of 0% - 100% vol. O2.
- Fast response: The response time (t90) is less than 15 seconds (from air to 100% O2).
- Good linearity: It has a linear output from 0% to 100% O2, which meets the requirements of ISO 80601 - 2 - 55.
- Stable baseline: The baseline offset is less than 200 μV.
- Temperature compensation: The temperature compensation is less than 2% O2 equivalent in the range of 0 °C to 40 °C.
- Wide operating temperature range: It can work in the temperature range of -20 °C to +50 °C.
- Wide operating pressure range: The operating pressure range is 0.5 bar to 2.0 bar.
- High - humidity resistance: It can operate in an environment with a humidity range of 0% to 99% rh (non - condensing).
- Low long - term output drift: In 100% O2, the long - term output drift is less than 5% signal loss per year.
Uses:
The MOX1 oxygen sensor is designed to monitor the partial pressure of oxygen in anesthesia (not including xenon), critical care, neonatal incubators, and general oxygen monitors.
| Output | 9 mv to 13 mv in 210 mbar O2 |
| Measurement Range | 0 mbar to 1500 mbar O2 |
| Response Time (t90) | < 15 s (air to 100% O2) |
| Baseline Offset | < 200 μV |
| Temperature Compensation | < 2% O2 equivalent (0 °C to 40 °C) |
| External Load Resistor | 10 kΩ minimum |
| Connector | 3 - pin Molex header (Molex 22 - 29 - 2031) |
| Housing Material | White ABS |
| Weight | 39 g (nominal) |
| Operating Temperature Range | - 20 °C to + 50 °C |
| Operating Pressure Range | 0.5 bar to 2.0 bar |
| Operating Humidity Range | 0% to 99% rh (non - condensing) |
| Long - term Output Drift in 100% O2 | < 5% signal loss/year |
| Recommended Storage Temperature | - 10 °C to + 40 °C |