8414776 Repetitive Exhalation Valve Of The Ventilator
Basic Information:
- Model: 8414776
- Product Type: Reusable Inhalation Valve / Inspiratory Valve (Ventilator Accessory)
- Application: Compatible with Drager Evita series ventilators (e.g., Evita 4, Evita XL, etc.)
- Purpose: Controls the flow of inhaled gas, located at the ventilator's inhalation port to regulate air delivery to the patient.
- Regulatory Approval: CE marked; complies with international medical device standards (e.g., ISO 13485).
Working Principle:
- Operates based on pressure differential sensing
- Opens when the ventilator delivers inspiratory gas (positive pressure), allowing gas to flow to the patient.
- Closes during exhalation to prevent backflow of exhaled gas into the ventilator circuit.
Material & Design:
- Material: Medical-grade plastic (biocompatible, corrosion-resistant), often with components like:
- Polycarbonate housing for durability.
- Silicone seals for airtightness and flexibility.
- Features:
- Reusable design with sterilization compatibility.
- Low dead space to minimize gas retention.
- Easy disassembly for cleaning and maintenance.
Compatibility:
- Specifically designed for Drager Evita ventilators; not compatible with other brands' systems.
Clinical Features:
- Airflow Control: Ensures precise delivery of tidal volume and inspiratory pressure.
- Hygiene & Safety: Biocompatible materials reduce the risk of patient infection.
- Ease of Use: Quick installation and removal for routine maintenance.
Warning & Maintenance:
- For medical use only: Must be operated by trained healthcare professionals.
- Sterilization Note: Follow Drager’s official guidelines to avoid material degradation (e.g., do not use harsh chemicals).
- Disposal: Discard immediately if physical damage (e.g., cracks, deformations) is observed.
Technical Parameters:
Parameter | Details |
Dimensions (Approx.) | Length: 12 cm, Diameter: 4 cm (model-specific; refer to product label) |
Weight | 60 g (approx.) |
Sterilization Methods | Autoclavable (134°C, 3 bar, 10 min); Ethylene oxide (EO) sterilizable. |
Reuse Cycle | Recommended for up to 20 uses or 90 days (follow clinical guidelines). |
Operating Temperature | 5°C–40°C (storage: -20°C–50°C) |
Humidity Resistance | ≤80% relative humidity (non-condensing) |