S6200 Anesthesia System For Adult And Pediatric
Application
The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room, Anesthesiology Depart- ment and other departments.
Professional design for adult, child and infant inhalation anesthesia and respiratory management, with advanced ventilation modes. Outstanding ergonomic design, it ranks high level in safty, stability and convenience as well as user experiences.
S6200 high-end model combine proven ventilation technology with the latest refinements in ergonomics and systems integration with an advanced, easy to use anesthesia table designed together with experienced experts to streamline your anesthesia workflow.
Trust point
- Patient Centered Ventilation: Precision in an anesthesia ventilator, from conventional ventilation to advanced modes and adapt to wide range patient.
- Safty design: Vaporizer with temperature, pressure, flow compensation and self-lock function. Real time pressure-time, flow-time loop oscillogram and high precision ETCO2, O2 concentration detection function included.
- Alarm: 3 level alarm system, visual and sound alarm information.
- Power: Built-in battery ensure 2-3 hours using when power failure.
- Separate design of electric circuit and gas circuit ensure long using life.
- Flexible configurations able to customize your requirements.
- Designed and manufactured by Superstar Medical with over 25 years experience in this area.
- Altitude compensation: Suitable for high/low altitude areas.
| Ventilation mode |
| IPPV, V-A/C, P-A/C, V-SIMV, P-SIMV, |
| PCV-PC, PCV-VG, PSV, SIGH, MANUAL |
| Ventilator parameter range |
| Flowmeter | O2(0.2 - 15L/min) N2O(0.2 - 12L/min) Air(0.2 - 15L/min) |
| Rapid oxygen supply | 25 L/min - 75 L/min |
| Tidal volume | 0, 10 mL - 1500 mL |
| Frequence(Freq) | 1 /min - 100 /min |
| I:E | 4:1 - 1:8 |
| PEEP | 0 cmH2O - 30 cmH2O |
| Pressure triggering sensitivity (PTr) | -20 cmH2O - 0 cmH2O (Based on PEEP) |
| Flow trigger sensitivity (FTr) | OFF, 0.5 L/min - 30 L/min |
| Pressure control (PC) | 5 cmH2O - 70 cmH2O |
| SIGH | 0 (off) 1/100 - 5/100 |
| Apnea ventilation | OFF, 5% - 60% |
| Pressure limit | 10 cmH2O - 100 cmH2O |
| Alarm and protection |
| The AC power failure alarm | Power failure or no connection |
| Internal battery backup low voltage alarm | < 10.2 ± 0.3 V |
| No tidal volume | ≤ 5 mL within 6 s |
| High oxygen concentration alarm | 19 % - 100 % |
| Low oxygen concentration alarm | 18 % - 99 % |
| High airway pressure alarm | 2 cmH2O - 100 cmH2O |
| Low airway pressure alarm | 0 cmH2O - 98 cmH2O |
| High minute volume alarm | 0.2 L/min,0.5 L/min - 100 L/min; |
| Low minute volume alarm | 0 L/min - 99 L/min |
| Continuous pressure alarm | (PEEP+1.5 kPa) over 16 s |
| Suffocation warning | 5 s - 60 s no spontaneous ventilation |
| The maximum limited pressure | ≤ 12.5 kPa |
| Fan error | Show on screen |
| Oxygen deficit | Show on screen |
| Monitoring parameter |
| Frequence(Freq) | 0 /min - 120 /min |
| Tidal volume (Vt) | 0 mL - 3000 mL |
| MV | 0 L/min - 100 L/min |
| Oxygen concentration | 15 % - 100 % |