| Material |
Grade |
316L Stainless Steel (UNS S31603, 1.4404) – Medical/Vacuum Melted Grade preferred. |
| |
Key Properties |
Excellent corrosion resistance, high biocompatibility (ISO 10993), good mechanical strength. |
| Capillary Tube |
Outer Diameter (OD) / Gauge |
27G to 34G (0.41 mm to 0.16 mm) for fine applications; 18G to 26G (1.27 mm to 0.46 mm) for standard. |
| |
Inner Diameter (ID) |
Proportionate to OD and wall thickness. Tightly controlled for consistent flow. |
| |
Wall Thickness |
Ultra-thin to standard wall. |
| |
Tube Length |
10 mm to 100 mm (customizable). |
| Tip & Orifice |
Bevel Type |
Standard Short Bevel or Long Taper Bevel for spray shaping. |
| |
Bevel Angle |
15°, 20°, 30° (customizable). Defines spray angle and penetration. |
| |
Orifice Type |
Open Bevel (for jet/stream) or Spray-Tip Ground (for fan/mist patterns). |
| |
Orifice Finish |
Mirror-polished and electropolished to prevent clogging and ensure smooth flow. |
| Hub/Connection |
Standard Connection |
Female Luer Lock (ISO 80369-7). |
| |
Alternative |
Luer Slip, or custom thread/fitting. |
| Performance |
Fluid Path |
Smooth, electropolished internal lumen (Ra < 0.2 µm). |
| |
Spray Pattern |
Linear Jet, Fine Mist, or Fan Spray (pattern depends on bevel/orifice grind). |
| |
Flow Rate Range |
From µL/min to mL/min, pressure-dependent. Can be characterized. |
| Marking |
Method |
Fiber Laser Marking. |
| |
Content |
Gauge (e.g., "30G"), Length (e.g., "13mm"), Part/Lot Number. |
| |
Durability |
Permanent, withstands sterilization. |
| Sterility & Compliance |
Sterilization (Single-Use) |
Gamma Irradiation or ETO. SAL 10⁻⁶. |
| |
Biocompatibility |
Complies with ISO 10993-1 (Cytotoxicity, Sensitization, Irritation). |
| |
Regulatory |
Manufactured under ISO 13485. Supports FDA/CE regulatory submissions. |
| |
Packaging |
Individual sterile blister pack within Tyvek®/plastic peel pouch. |