Technology Parameters of Syringe Pump

| Main features |
| Reliable and Safe: |
| Advanced safety design able to stop infusion and alarm even if the CPU damage,the medical staff can use it safely. |
| Applicable syringe |
| 5ml,10ml,20ml,30ml,50/60ml |
| The maximum flow rate range |
| 5ml syringe: 0.1ml/h~200ml/h |
| 10ml syringe: 1ml/h~300ml/h |
| 20ml syringe: 1ml/h~400ml/h |
| 30ml syringe: 1ml/h~600ml/h |
| 50/60ml syringe: 1ml/h~1200ml/h |
| Precision |
| Better than ±2%; |
| (but 5ml syringe and <1ml/h: Better than ±8%) |

| Parameter range |
| 1, Flow-rate: 0.1~1200.0 ml/h, 0.1ml/h step; |
| 2, Dose: 0.1~999.9 mg, 0.1mg step; |
| 3, mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step; |
| 4, Body-weight: 1~300 kg, 1kg step; |
| 5, KVO rate: 0.1~5.0 ml/h, 0.1 ml/h step; |
| 6, Bolus rate: 0.1~1200.0 ml/h, 0.1 ml/h step; |
| 7, Bolus volume: 0.1~99.9 ml/h, 0.1ml step |
| Alarm |
| 1, Occlusion,0~175KPa, 8 grades of sensitivity; |
| 2, Pressure sensor is abnormal; |
| 3, Push rod fall off; |
| 4, The drug running out; |
| 5, The injection will finish; |
| 6, The motor rotation fault; |
| 7, Overtime pause; |
| 8, Low battery; |
| 9, Battery failure; |
| 10, Battery temperature is too high; |
| 11, Mains missing tips; |
| 12, MCU fault, and stop injection. |
| Safety classification |
| Class I; Internal power supply, CF type, IPX4 equipment |

| Power supply |
| AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery |
| Power consumption:20VA |
| Working time of battery: 10h@5ml/h |
| Weight and size |
| 37cm(w)*24cm(h)*21cm(t); about 2.1kg |
| Optional accessories |
| Wireless communicating module. |
| Chinese Certificate |
| yes |