This calibration device is designed in accordance with ASTM D5207-03 and related IEC, UL, GB standards for measuring flame temperatures in combustion test equipment.
Position the copper block 10±1mm above the needle flame burner. Ignite the flame; the time for the copper block temperature to rise from 100°C±2°C to 700°C±3°C must be 44±2 seconds to validate the flame for testing. Perform three measurements, allowing the copper block to cool naturally to <50°C between tests.
Model | 6014-HD |
Thermocouple | Armored K-type (NiCr/NiAl) fine-wire thermocouple with a 0.5mm diameter sheath |
Timer | Adjustable range: 0.01-99.99 seconds |
Copper Block Specifications | |
Diameter | Ø5.50±0.01mm |
Length | 8.80±0.02mm |
Bevel Angle | 45±0.5° |
Mass (Pre-drilling) | 1.76g±0.01g |
Material | High-conductivity electrolytic copper |
Surface Finish | Fully polished |