This calibration device is designed in accordance with ASTM D5207-03 and related IEC, UL, GB standards for measuring flame temperatures in combustion test equipment.
Position the copper block 10±1mm above the needle flame burner. Ignite the flame; the time for the copper block temperature to rise from 100°C±2°C to 700°C±3°C must be 44±2 seconds to validate the flame for testing. Perform three measurements, allowing the copper block to cool naturally to <50°C between tests.
| Model | 6014-HD |
| Thermocouple | Armored K-type (NiCr/NiAl) fine-wire thermocouple with a 0.5mm diameter sheath |
| Timer | Adjustable range: 0.01-99.99 seconds |
| Copper Block Specifications | |
| Diameter | Ø5.50±0.01mm |
| Length | 8.80±0.02mm |
| Bevel Angle | 45±0.5° |
| Mass (Pre-drilling) | 1.76g±0.01g |
| Material | High-conductivity electrolytic copper |
| Surface Finish | Fully polished |




Why choose us?
1.Design & Production : Solution Consultation for each order
Recommend the most cost -performance machine
Research & Develop the suitable one for Special- designed Model.
2.Shipping & Delivery: One-Stop Service
Production- Deliver to your door - Installation & Training- Technical Support
3.Warranty
1 year warranty with lifetime technology support
4.After-Sale Service
React within 2 Hours
Video inspection & on-site service combined
FAQ
Q: How about your warranty?
Q: What is the order process?
Q: Can we visit your factory before order?
Q: Do you accept customization service?