Insulation Oil Dielectric Dissipation Factor Volume Resistivity Tester
I. Introduction:
Automatic Dielectric Dissipation Factor Tester is a high-precision instrument, which can be applied to measure dielectric loss angle and Volume Resistivity of insulating oil etc.
II. Specifications:
| Condition of Use | -5℃-40℃ RH<80% | 
| Power Source | AC 220V±10% Without Frequency Limit | 
| AC High Voltage Output | 400V-2200V ±2% Every 100V 50VA | 
| DC High Voltage Output | 200V-600V ±2% | 
| Temperature Control Induction Stove | Maximal Power 500W | 
| Temperature Control Range | <100℃ | 
| Temperature Measuring Error | ±0.5℃ | 
| Control Temperature Error | 0.1℃ | 
| Control Temperature Time | Room Temperature 90℃ <20min | 
Measuring Range  |  			 			 tgδ Without Limit Cx 15PF-300PF Rx 10M-10T  |  		
| Resolution |  			 tgδ 0.001% Cx 0.01pF Rx 0.01  |  		
| Precision |  			 delta tgδ:±(Reading*0.5%+0.020%) delta Cx :±(Reading*0.5%+0.5PF) delta Rx :±Reading * 10%  |  		
| Relative Dielectric Loss Constant | Εr Calculated Automatically | 
| Volume Resistivity | ρ Calculated Automatically | 
| Dimension | 450(L)×310(W)×360(H) | 
| Net Weight | 21KG |  		
Oil Cup of Dielectric Dissipation Factor Volume Resistivity Meter