Hematology Analyzer Reagents Excellent Performance Hematology Reagent For Nihon Khoden 3 Part Diff Analyzers
Intended Use
To be used for blood cell counting, sizing, and WBC differentiation, hemoglobin determination on Nihon Khoden 3 part diff hematology analyzers. This product is for In Vitro Diagnostic Use only.
Applicable Instument
For use with Nihon Kohden MEK-6318,MEK-6400,MEK-6700, MEK-6410,MEK-6420 hematology analyzers.
Procedures
- Remove all the reagent aspiration tubes from the currently used reagent boxes.
- Wipe the tube’s surface with a lent free tissue, and then rinse with distilled water. Put the tubes into Labnovation reagents boxes or bottles.
- Run at least 50 backgrounds. Check for the acceptable background values. At first background counts can be seen, that should lower to normal after a few runs.
- Run all three levels of a dedicated commercial control. Check for the acceptable values.
- Recalibrate the unit if necessary.
Precaution
- Re-calibration is recommended when changing the origin of the reagents.
- It is not recommended using different brands of solutions together on one unit of analyzer.
- Do not ingest. Avoid contact with skin and eyes.
- Skin contact: Flush with plenty of water.
- Eyes contact: Flush with plenty of water. Obtain immediate medical attention.
- This kit is for professionals and in vitro diagnostic use only.

Specification
41-10-020 | Diluent NK | 20L |
41-30-A05 | Lyse NK | 500ML X3 |
41-40-005R | Rinse NK | 5 L |
L-0401005 | Cleaner-C | 500ML |
KIT COMPOSITION
DILUENT NK | NaCl≤9.4%, Buffer≤1.6%, Anti-Microbial Agent≤1.0% |
CLEANER-C | Sodium Hypochlorite 5% |
LYSE NK | Quaternary Ammonium Salt<10%, Sodium Lauryl Sulfate<1.5% |
RINSE NK | Buffer<0.9%, Sufurctant<1.0% |
Storage
Store the product at 5-35℃ in a dust free place with good ventilation. Temperature outside the range may damage the product.
The open vial life of the product is 30/60 days. Unused reagents should be discarded after 30/60 days. Do not mix residual with a new carton.
Advantage
- One source for all brands of analyzer
- Reagents for both 3-parts and 5-parts
- Ready-to-use final products and bulk OEM reagents
- ISQ13485:2003 and ISO9001:2008 certified
- CE marked, FDA cleared
- Well-established brand used by thousands of lab in 95 countries
- High quality, high performance and high fidelity of correlations with original brand