Deltapilot M FMB50 Hydrostatic Level Transmitter
Compact pressure sensor with Contite measuring cell for hydrostatic level measurement in liquid and paste-like media
Product Overview
The compact Deltapilot FMB50 transmitter with condensate-resistant Contite measuring cell is designed for level measurement in open or closed containers, unaffected by foam formation. Suitable for SIL2 safety systems.
Key Features
- Hermetically sealed Contite measuring cell with condensate-resistance
- High reference accuracy: ±0.2%, optionally ±0.1%
- Minimal temperature effects
- Modular concept for easy replacement of display or electronics
- Seamless system integration (HART/PA/FF)
- Easy menu-guided commissioning via local display
Technical Specifications
| Parameter | Specification |
| Measuring Principle | Hydrostatic pressure |
| Supply Voltage | 4...20 mA HART: 10.5...45 VDC (Non Ex) Ex ia: 10.5...30 VDC PROFIBUS PA: 9...32 VDC (Non Ex) FOUNDATION Fieldbus: 9...32 VDC (Non Ex) |
| Reference Accuracy | Standard 0.2% Optional 0.1% |
| Long Term Stability | 0.05% of URL/year |
| Process Temperature | -10°C...100°C (14°F...212°F) |
| Ambient Temperature | -40°C...85°C (-40°F...185°F) |
| Measuring Cell | 100 mbar...10 bar (1.5 psi...150 psi) |
| Smallest Calibratable Span | 10 mbar (1.45 psi) |
| Vacuum Resistance | 0 mbar abs. |
| Max. Turn Down | 100:1 |
| Max. Overpressure Limit | 40 bar (600 psi) |
| Communication | 4...20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus, IO-Link |
Process Connections & Materials
| Component | Specification |
| Process Connection | Thread: G1 1/2, MNPT1 1/2 Flange: DN40...DN100, ASME 1 1/2"...4", JIS 10K |
| Hygienic Connections | Tri-Clamp, DIN11851, DIN11864-1, NEUMO, Varivent N, SMS, DRD |
| Process Membrane Material | 316L, AlloyC, Gold-Rhodium |
| Gasket Material | None, measuring cell welded |
| Fill Fluid | Inert oil, Synthetic oil |
| Housing Material | 316L, Die-cast aluminum |
Certifications & Approvals
- Certificates: ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex
- Safety: Overfill protection WHG, SIL
- Design: EN10204-3.1, NACE MR0103
- Hygienic: 3A, EHEDG, CoC ASME-BPE
- Marine: GL/ABS/LR/BV/DNV
- Drinking Water: KTW/NSF/ACS
Applications
Ideal for process and hygienic applications in various industries including chemical, pharmaceutical, food & beverage, and water treatment for level measurement in liquids and paste-like media.
Company Information
Frequently Asked Questions
| Question | Answer |
| Manufacturer or Trade Company? | Manufacturer with over 10 years experience in control systems and instrumentation. OEM & ODM available. |
| Samples & MOQ? | Sample orders supported. |
| Lead Time? | Generally 5-20 days (based on quantity), most models in stock. |
| Loading Port? | Guangzhou Port, China. |
| Payment Terms? | T/T, PayPal, Western Union. |
Our Advantages
- Best possible prices and delivery time
- Long time experience in the field
- Well trained staff
- Own warehouse facilities
- Best relationships with branch office of manufacture
- Flexible and individual service
- Cheap and safe forwarder for shipments (air)