PTM100 Portable VOC Analyzer With Flame Ionization Detector FID
Overview
The PTM100 volatile organic gas analyzer features both flame ionization (FID) and photoionization (PID) detectors that operate simultaneously in an explosion-proof design. This instrument is ideal for LDAR detection, leakage detection in oil and gas collection systems, VOC detection on liquid surfaces, rapid soil pollutant screening, and comprehensive regional surveys.
Features & Benefits
- Dual detection with hydrogen flame ionization (standard) and photoionization
- Digital hydrogen pressure display with intelligent remaining working time reminder
- Battery power monitoring with remaining time indicator
- FID automatic ignition with real-time flame temperature monitoring
- Built-in database of over 200 volatile organic gases
- Explosion-proof host (Ex db ia IIC T4 Gb)
- 6.5-inch touch screen handheld operator with Bluetooth and WIFI support
- Explosion-proof handheld operator (Ex ib IIC T4 Gb/Ex ibD 21 T130℃)
Application
This analyzer is used for gas analysis instrument testing and calibration, reactor gas supply, sample gas debugging, and gas mixture synthesis. It can pre-set output gas concentration and duration to automatically complete gas matching experiments. Applications include testing methods such as HJ1006-2018, HJ 604-2017, HJ 1261-2022, and CO/SO2 matrix testing in HJ57-2017.
Product Parameters
| Main parameters |
Parameter range |
| Measuring range |
FID: 0~100,000 ppm PID: 0~2,000 ppm (multiple ranges available, optional configuration) |
| Method detection limit |
FID: 0.5 ppm methane PID: 0.5 ppm isobutylene |
| Accuracy |
±5% of reading or ±1 ppm, whichever is greater |
| Repeatability |
FID: 2% at 500 ppm methane PID: 1% at 100 ppm isobutylene |
| Response time (T90) |
FID: ≤3.5 s (10000ppm methane) PID: ≤3.5 s (500ppm isobutylene) |
| Sampling flow |
800 ml/min (adjustable) |
| Battery life |
≥10 hours (0℃) |
| Hydrogen endurance |
≥15 hours |
| Power adapter |
DC8.4 V 1A |
| Work environment |
-30~45°C, 10%~95% RH |
| Gas interface |
1/4 inch |
| Host weight |
3.5 KG |