PH010 Newest Water Quality Measurement Tool PH Pen
      OVERVIEW
   
  Based on the old model PH water test pen ,the new model changed the switch, add automatic Startup & Shutdown function, and program optimization, etc. For example, switch from the previous horizontal push and pull, make it directly up and down button way, more conveniently to use for customers
  
  
    CALIBRATION
    	- Put the electrode into a mixed phosphate standard buffer of pH 6.86 (25 ° C) and stir gently until value is stable.
  	- Adjust the calibration potentiometer with a small screwdriver until the display value matches the pH value of the standard buffer solution at nomal temperature.
  	- The electrode is inserted into a standard buffer solution of borax with potassium hydrogen phosphate at pH4.01.
  	- The display value should be within the allowable error range when compared to the pH of the buffer solution.
  
       
  SPECIFICATION
    
   	 		 			| Name |  			PH meter |  			Power Supply |  			2*1.5V (LR44 Cell) |  			Dimension |  			130*29*20mm |  		
 		 			| Range |  			0.0-14.0PH |  			Temperature compensation |  			0℃ -50℃ |  			Weight |  			51g |  		
 		 			| Resolution |  			0.1PH |  			Operating Temperature |  			0℃ -80℃ |  			Color |  			white , or customer required colors |  		
 		 			| Accuracy |  			±0.1PH |  			Calibration |  			 			 Two-Point   			Auto-Calibration  			 |  			Service |  			OEM/ODM |  		
 	 
  
     