The 5-21R High-Speed Refrigerated Centrifuge is designed for modern laboratories requiring fast, precise, and reliable sample separation. Equipped with high-speed rotors and advanced refrigeration technology, it maintains sample integrity even under prolonged operation. Its intelligent safety system prevents accidents and protects valuable samples. Perfect for clinical laboratories, research institutes, pharmaceutical labs, and biotech companies.
| Parameter | Specification |
|---|---|
| Maximum speed | 21000r/min |
| Maximum relative centrifugal force | 53628xg |
| Maximum capacity | 6×500ml |
| Rotational speed accuracy | ±10r/min |
| Timing range | 0~99min59sec |
| Temperature range | -20°C to 40°C |
| Total power | 4.5kw |
| Power supply | Ac220V 50Hz 35A |
| Noise of the whole machine | ≤65dB(A) |
| Net weight | 190kg |
| Display screen | Color |
| Rotor name | Maximum speed | Maximum relative centrifugal force (xg) | Angle |
|---|---|---|---|
| 30*1.5ml | 21000rpm | 53628xg | 45° |
| 16*10ml | 21000rpm | 53000xg | 30° |
| 6*50ml | 20000rpm | 43370xg | 30° |
| 8*50ml | 18000rpm | 39945xg | 28° |
| 8*100ml | 18000rpm | 44617xg | 30° |
| 6*250ml | 14000rpm | 36260xg | 30° |
| 6*500ml | 8000rpm | 11680xg | 30° |
1. The power supply requirements are 3×6mm² lines, single-phase AC 50Hz power supply, the voltage is 220V, the voltage requirements are relatively stable, the fluctuation requirements are between ±10%, and the rated current is 30A. The power supply voltage should be consistent with the voltage index on the nameplate of the instrument. If the difference between the power supply voltage and the rated voltage is more than 10%, the power supply does not meet the installation requirements. (If the power supply voltage cannot meet the requirements, it is recommended to equip the machine with a fully automatic electronic AC voltage regulator purifier).
2. The user shall provide 16A power plug and socket or circuit breaker, and the position of the socket or circuit breaker shall not be more than 1 meter away from the position of the centrifuge.
3. In order to ensure safety, the centrifuge should be connected to a remote emergency switch (the switch is preferably outside the room where the centrifuge is located or near the exit), so that the connection between the centrifuge and the main power supply can be cut off in the event of a failure.
4. Please make sure that the socket you want to use is properly wired and grounded. Never use a 3-wire to 2-wire adapter; Two-wire extension wires or two-wire grounding multi-socket terminal blocks should never be used.