Thermal Abuse Test Chamber for Battery Cell
1. IEC 62133-2017
2. UL 1642
3. GB 31241-2014
| Model
| BE-8103 |
|---|---|
| Structure
| All-in-one structure, the door uses explosion-proof lock + explosion-proof chain |
|
| Normal temperature to 150°C (Adjustable) |
| Temperature Range
| |
| Controlling Accuracy
| ±0.5℃ |
| Temperature Deviation
| ±2.0℃, No load |
| Heating Rate
| RT~150°C (Linear 5°C/min) |
| Control Display
| PLC touch type (7 inch) + remote control |
| Working Cabinet Dimension
| W600XD500XH600 mm |
| External Dimension
| W940XD780XH1620 mm |
| Inner Chamber Material
| SUS304 Stainless steel with mirror-finish |
| External Chamber Material
| SECC steel plate, fine powder paint coating (thickness 1.5mm) |
| Bottom
| With universal wheel |
| Observation Window
| 390 X 360mm (20mm thick explosion-proof glass, explosion-proof mesh) |
| Insulation Material
| Imported high-density glass wool, good heat insulation, to ensure maximum temperature insulation, tight sealing materials for high temperature resistant silicone, withstanding high temperatures over 300 degrees for a long time |
| Terminal Hole
| Diameter 50mm Terminal hole on the left and right sides |
| Power Supply
| 220V / 50HZ / 4.0KW |
| USB Interface
| With USB interface, the test data can be exported |
| Auxiliary Function
| Explosion-proof pressure relief device, smoke exhaust device, explosion-proof lighting device, electronic door lock system, safety stop switch |

A: Relevant standard test requirements
Single battery safety test
The heating test is carried out as follows:
1. The single battery is charged according to the method of 6.1.3
2. Put the single battery into the temperature chamber.
3. For lithium-ion batteries, the temperature chamber is increased from room temperature to 130 ° C ± 2 ° C at a rate of 5 ° C / min, and the heating is stopped after maintaining the temperature for 30 minutes.
4. For nickel metal cyanide batteries, the temperature chamber is raised from room temperature to 85 ° C ± 2 ° C at a rate of 5 ° C / min, and the heating is stopped after maintaining this temperature for 2h.
5. Observe for 1h.
B: Battery module safety test
1. The battery module battery shall be charged according to the method in 6.1.4.
2. For lithium-ion batteries, the temperature chamber is raised from room temperature to 130 ° C ± 2 ° C at a rate of 5 ° C / min, and the heating is stopped after maintaining this temperature for 30 minutes.
3. For nickel metal cyanide batteries, the temperature chamber is raised to 85 ° C ± 2 ° C at room temperature at a rate of 5 ° C / min, and the heating is stopped after maintaining this temperature for 2h.
4. Observe for 1h.