This is a multifunctional digital power meter for measuring the power of single - phase (1P/2W) AC and DC power supplies. The test frequency range of this product can reach DC 0.5Hz - 100kHz. It uses a 5" IPS LCD display screen, has a 5 - digit measurement display, enables simultaneous measurement of multiple parameters, and possesses high - precision measurement capability. At the same time, it has functions such as waveform display (voltage/current/power), integral measurement function, measurement and analysis of each harmonic (measurement at 50/60Hz meets the requirements of IEC61000 - 4 - 7), external sensor input terminals, and various communication interfaces to assist users in achieving clear, convenient and accurate power measurement.
|
Input indicators |
|||
|
Item |
Voltage |
Current |
|
|
Input type |
Floating input via resistor divider |
Direct input or sensor input |
|
|
Input impedance |
Approx. 2MΩ |
Direct input:: Approx. 500mΩ (200mA and below range) Approx. 50mΩ (0.5A and above range) Sensor input: about 100KΩ |
|
|
Measurement range (RMS) |
15V, 30V, 60V, 150V, 300V, 600V |
Direct input: 5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 0.5A, 1A, 2A, 5A, 10A, 20A Sensor input: 50mV, 100mV, 200mV, 500mV, 1V, 2V, 5V, 10V |
|
|
Maximum continuous input |
1.5kV Peak or 1kV RMS, whichever is smaller |
30A Peak or 20A RMS, whichever is smaller (200mA and below) 100A Peak or 30A RMS, whichever is smaller (0.5A and above) |
|
|
Voltage and current accuracy |
|||
|
Item |
Specification |
||
|
DC |
±(0.1% reading + 0.2% range) |
||
|
0.5Hz ≤ f < 45Hz |
±(0.1% reading + 0.2% range) |
||
|
45Hz ≤ f ≤ 66Hz |
±(0.1% reading + 0.05% range) |
||
|
66Hz < f ≤ 1kHz |
±(0.1% reading + 0.2% range) |
||
|
1kHz < f ≤ 10kHz |
±[(0.07×f)% reading + 0.3% range] |
||
|
Temperature coefficient |
5 to 18°C or 28 to 40°C: Add ±0.03% of reading. |
||
|
When the power filter is on |
<45Hz: add 0.5% of the reading; 45~66Hz: add 0.2% of the count |
||
|
Crest Factor Impact |
The range error when the crest factor is 6 is twice the range error when the crest factor is 3. |
||
|
Effect of temperature change after zero calibration or range change |
Add 0.02%/°C of range to DC voltage accuracy. Add the following values to DC current accuracy: 200mA and below range: 5uA/°C; 0.5A and above range: 500uA/°C. External sensor input: 1mV/°C. |
||
|
Active power measurement accuracy |
|||
|
Item |
Specification |
||
|
Measurement conditions |
Temperature: 23±5℃ Humidity: 30~75%RH (no condensation) After 30 minutes of preheating Input waveform: stable DC or sine wave Common mode voltage: 0V Crest factor: 3 Power filter: OFF Frequency filter: On to measure ≤200Hz voltage or current Zero calibration after measuring range change |
||
|
DC |
±(0.1% reading + 0.2% range) |
||
|
0.5Hz ≤ f < 45Hz |
±(0.3% reading + 0.2% range) |
||
|
45Hz ≤ f ≤ 66Hz |
±(0.1% reading + 0.05% range) |
||
|
66Hz < f ≤ 1kHz |
±(0.2% reading + 0.2% range) |
||
|
1kHz < f ≤ 10kHz |
±(0.1% reading + 0.3% range) ±[0.067×(f-1)% reading] |
||
|
10kHz < f ≤ 100kHz |
±(0.5% reading + 0.5% range) ±[0.09×(f-10)% reading] |
||
|
Temperature coefficient |
5~18℃ or 28~40℃: add ±0.03% of reading.
|
||
|
When the power filter is on |
<45Hz: add 1% of reading; 45~66Hz: add 0.3% of count. |
||
|
Crest Factor Impact |
The range error when the peak factor is 6 is twice the range error when the peak factor is 3. |
||