Analytical Industries PSR11-915-4 medical oxygen sensor for datex ohmeda
Specification:
 
 All characteristics are based on conditions at 25°C, 50% RH and 1013 hPa 
 Electrical Connector: modular jack 6P4C 
 Measurement Range: 0 to 100 Vol.% 
 Nominal Sensor Life: > 500, 000 Vol.% h 
 Expected Operating Life: < 3 years @ ambient air 
 Output Signal: 8.0 to 12 mV @ dry ambient air 
 Response Time t90 : < 12 s 
 Drift: < 1% volume O2 / month @ air 
 Operating Temperature: 0 to 40°C 
 Pressure Range: 750 to 1250 hPa 
 Linearity Error: ≤ 3% @ 100% O2 applied for 5 min 
 Zero Offset Voltage: < 200 µV in 100% N2, applied for 5 min 
 Repeatability: ± 1% volume O2 @ 100% O2, applied for 5 min 
 Influence of Humidity: -0.03% rel. O2 reading per % RH 
 Recommended Load Resistor: ≥ 10 kOhm 
 Temperature Compensation: NTC 
 Interferences: according to DIN EN 12598 and ISO 7767 
 
 Temperature Range: recommended: 5 to 30°C 
 maximum: - 20 to 50°C 
 Humidity: up to 100% RH 
 Shelf Life: < 6 months 
 Weight: approximately 52 g 
 Material in contact with media: PA, PPS, PTFE, stainless steel 
 Warranty Period: 12 months 
 Housing Colour: black
Replacement:
Mainly used in Europe and the United States MAZUI machine AESPIRE, AESTIVA7100, AESTIVA7900 AESPIRE7900, AVANCE ADU AYSIS.
 OHMEDA oxygen battery, OHMEDAMAZUI oxygen battery