| Area Classification | General Purpose Class I Div II (without the use of IS Barrier) Class I Div I (with the use of IS Barrier) |
|---|---|
| Enclosure Material | Plastic (Polycarbonate) |
| Housing Rating | IP66/ NEMA 4A/ CSA 3S |
| Power Supply | 2-wire 24 VDC Loop Powered |
| Output Signal | Bi-directional HART digital communication, superimposed on mA (4-20 mA) signal Optional FF (FOUNDATION Fieldbus) and PF (Profibus) communication |
| Intrinsically safe type | ATEX, IECEx, FM, CSA and NEPSI, and nonincendive of FM and CSA. |
Built-in calibration standard table, with temperature compensation function and stability check function, realizes high-precision pH measurement.
In addition to pH measurements, temperature and redox potential (ORP) can also be measured.
As a diagnostic function of the detector, impedance, asymmetric potential, slope, etc. are constantly measured to continuously diagnose the state of the detector such as dirt/damage of the electrode, disconnection, and lowering of the measuring liquid level. In addition, the latest five calibration results are stored to predict future maintenance and calibration timing.

| Input Range | pH: -2 to 16 pH (with option /K: 0 to 14 pH) ORP: -1500 to 1500 mV rH: 0 to 100 rH Temperature: Pt1000, Pt100, 6.8k, PTC10k, 3k Balco, PTC500: -30 to 140 ºC, NTC 8k55: -10 to 120 ºC | |
|---|---|---|
| Performance (The specifications are expressed with simulated inputs.) | pH | Linearity: ±0.01 pH, Repeatability: ±0.01 pH, Accuracy: ±0.01 pH |
| ORP | Linearity: ±1 mV, Repeatability: ±1 mV, Accuracy: ±1 mV | |
| Temperature | with Pt1000, PTC500, 6.8k, PTC10k, NTC 8k55, 3k Balco Linearity: ±0.3 ºC, Repeatability: ±0.1 ºC, Accuracy: ±0.3 ºC | |
| with Pt100 Linearity: ±0.4 ºC, Repeatability: ±0.1 ºC, Accuracy: ±0.4 ºC | ||
Built-in temperature compensation function and calibration standard table enable monitoring of the contamination of the cell for higher accuracy conductivity measurement.
Highly versatile, cell constants from 0.005 to 50 cm-1, two-way and four-way detectors are available.
It also supports five types of temperature detectors for accurate temperature compensation.
Performs pure water compensation in the analysis of demineralized water, steam, condensate, and boiler water. For monitoring of pure water in the pharmaceutical industry, the process liquid quality can be monitored according to the water quality standards of the USP item <645> (refer to item 645 of USP23 Article) guidance (USP).

| Input Specification | Two or four electrodes measurement with square wave excitation, using max 60m (200ft) cable (WU40/WF10) and cell constants from 0.005 to 50.0 cm-1 | |
|---|---|---|
| Display range | Conductivity | min. 0.01 µS/cm, max. 2000 mS/cm (max 90% zero suppression) |
| Resistivity | min. 0.001 kΩ x cm, max. 1000 MΩ x cm (max 90% zero suppression) | |
| Temperature | min. span 25 ºC, max. span 270 ºC | |
| Performance (Accuracy) (The specifications are expressed with simulated inputs.) | Conductivity | 2 µS x K cm-1 to 200 mS x K cm-1: Accuracy: ±0.5%F.S. 1 µS x K cm-1 to 2 µS x K cm-1: Accuracy: ±1%F.S. |
| Resistivity | 0.005kΩ / K cm-1 to 0.5MΩ /K cm-1: Accuracy: ±0.5%F.S. 0.5MΩ / K cm-1 to 1MΩ /K cm-1: Accuracy: ±1%F.S. | |
| Temperature | with Pt1000, Pb36, Ni100: Accuracy: ±0.3 ºC with Pt100, NTC 8k55: Accuracy: ±0.4 ºC | |
| Temperature compensation | NaCl table: ±1 % Matrix: ±3 % | |
| Temperature | 90 % (< 2 decades) in 7 seconds | |