Battery Testing Aging Test Chamber High Temperature Aging Test Chamber
The Aging Test Chamber is suitable for the whole machine device, parts, materials for high temperature accelerated aging test through the test of the equipment can be checked or evaluated product defects, to quickly find out the problem, provide effective means to solve the problem,fully improve production efficiency and product quality.
Feature:
- Test Space 1000L
- Internal size 1000mm×1000mm×1000mm(W×D×H)
- External size 1690mmx1280mmx1850mm (W×D×H)
- Power supply AC380V±10%,50Hz±1,60A, 3/N/PE, earth resistance < 4Ω
- Weight About 300kg
- Max. Currency 18.5A(Average operating currency 9A)
- Max. Power 12.5KW(Average operating power 6KW)
- Installation Space Work station ground should be flat, good ventilation; no flammable article, corrosive gas or dust
- No strong electromagnetic radiation source nearby
- Close to drainage system ( within 2 meters to refrigeration unit,for condensate water)
- The ground bearing capacity not less than 8000kg/m2
- Prepare proper power supply and water supply
- The ground bearing capacity not less than 8000kg/m2
- Prepare proper power supply and water supply
- If need water or gas supply, should be equipped with in advance
- Ambient Temp: 5℃~25℃suggested, 25℃~35℃ function well
- Ambient Humid:≤85%RH
- Ambient Pressure:86KPa~106KPa
- Installation space assurances, as following
|
Installation Space:
- Work station ground should be flat, good ventilation; no flammable article, corrosive gas or dust
|
- No strong electromagnetic radiation source nearby
|
- Close to drainage system ( within 2 meters to refrigeration unit,for condensate water)
|
- The ground bearing capacity not less than 8000kg/m2
|
- Prepare proper power supply and water supply
|
- If need water or gas supply, should be equipped with in advance
|
- Ambient Temp: 5℃~25℃suggested, 25℃~35℃ function well
|
|
- Ambient Pressure:86KPa~106KPa
|
- Installation space assurances, as following
 |
Specifications:
Model: | EWG2006 | EWG3006 | EWG2010 | EWG3010 |
Test Space Volume(L) | 1000 | 600 |
Temp.Range(℃) | R.T+20~200 | R.T+20~300 | R.T+20~200 | R.T+20~300 |
Temp.Constancy(℃) | ±0.1~±0.8 |
Temp.Homogeneity(℃) | ±0.5~±2.0(RT+20℃~+200℃),±1.5~±3.5(200℃~+300℃) |
Temp.Deviation(℃) | ±0.5~±2.0(RT+20℃~+200℃),±0.5~±2.0(200℃~+300℃) |
Heat Up Time (min) | 50(RT→+200℃),90(RT→+300℃) |
Test Space Dimension(mm) | 850W×700D×1000H | 1000W×1000D×1000H |
External Dimension(mm) | 1250W×980D×1850H | 1690W×1280D×1850H |
Power | AC380V±10%,50HZ,3/N/PE |
Aging Rooms | EWG25 | EWG45 | EWG65 | EWG85 | EWG100 | EWG120 |
Test Space Volume(m³) | 25 | 45 | 65 | 85 | 100 | 120 |
Temp.Range(℃) | RT+10~85 |
Temp.Constancy(℃) | ±0.5~±1.0 |
Temp.Homogeneity(℃) | ±1.5~±3.0 |
Temp.Deviation(℃) | ±1.5~±3.0 |
HeatingRate(℃/min) | 2℃/min on average |
Power | AC380V±10%,50HZ,3/N/PE |
Battery Test Chamber Options Accessories • Independent sensors for specimen protection(NE60519-2.1993) • 1 Entry port Φ100mm • Nitrogen gas auxiliary device • Spare parts package • E-management and cyber-software • Video monitor system • APP for mobility management | Battery Test Chamber Standard Version • 1 Entry port Φ50mm • 1 Silicon stopper for entry port • 2 Stainless steel shelves |