High And Low Temperature Test Chamber Two Sides Door Open For Oil
OEM and ODM are welcome
More detail information ,pls kindly contact :anna@komegtech.com Mobile no :+86-18098281986 or Wechat:+86-18098281986
|  			 Main technical parameters ( water-cooled, temperature +25℃, no load) 
  |  		|
| 1.Temperature range |  			 -70℃ ~ +180℃(Different temperature range per customer requirements) 
  |  		
| 2.Temperature deviation | ±2.0℃ | 
| 3.Temperature fluctuation | ≦1.0℃ | 
| 4.Temperature uniformity | ≦±2.0℃ | 
|  			 5.Heating up rate&Cooling down rate 
  |  			 			 from 25℃ to +150℃ whole process average≧3℃/min Load 50 kg aluminum alloy 
 from +25℃ to -40℃whole process average ≧2℃/min Load 50kg aluminum alloy  |  		
|  			 The temperature performance test is based on the relevant provisions of the IEC 60068 -3 
 standard; the sensor is placed at the unit air outlet.  |  		|
|  			 Chamber structure 
  |  		|
|  			 The test chamber adopts the overall structure, the upper part is the working chamber, the rear side is 
 the refrigeration unit, the lower right space (≧700mm) is vacant to install the hydraulic part device, the 
 right is the electric control cabinet, the control panel is located on the right front side; the width direction 
 opens the door , the right side is the air duct system; the air supply motor, the centrifugal fan blade, the 
 heater, and the evaporator. 
  |  		|
| 1.Chamber size |  			 Inner size: W 1800 × H 1200 × D 1200 mm 
 Outer size :W 2935 × H 2170 × D 1450 mm approx  |  		
| 2.Insulation box |  			 Outer wall material: cold-rolled steel plate electrostatic double-sided spray, color ivory white; 
 Inner wall material: SUS304# matte stainless steel plate, thickness 1.5mm 
 Insulation material: rigid polyurethane foam insulation layer + glass fiber.  |  		
| 3.Structural strength |  			 bottom bearing capacity: ≧150kg 
 The bottom of the workroom has a threaded port for mounting the work shelf  |  		
Standards Implement and Met
● GB-2423.1-2008(IEC68-2-1)Test A: Low Temperature Test.
● GB-2423.2-2008(IEC68-2-2)Test B: High Temperature Test
● GJB360.8-2009(MIL-STD.202F) High Temperature Life Test
● GJBl50.3-2009(MIL-STD-810D)High Temperature Test
● GJBl50.4-2009(MIL-STD-810D) Low Temperature Test
● GB2423.3-2008 (IEC68-2-3) Test Ca: Constant Hot and Humid Test
● GB2423.4-2008 (IEC68-2-30) test Db: Alternating Hot and Humid Test
● GJBl50.9-2009 (MIL-STD-810D) Damp heat test method
Photos show