1. Product Overview
The A2L refrigerant hot surface ignition temperature tester determines the lowest temperature at which A2L refrigerant can be ignited by a high-temperature hot surface under specific conditions; this is known as the hot surface auto-ignition temperature. This data is crucial for assessing the potential fire risk when refrigerant leaks and encounters hot surfaces such as electrical components or overheated pipes.
2. Technical Parameters
2.1 Test Chamber: a pressure-resistant and high-temperature-resistant glass tube .
Glass Tube Material: Borosilicate .
Glass Tube Dimensions: Diameter 70mm±10mm, Height 230mm±10mm
2.2 Thermal Surface Assembly
Hot Surface Assembly: 304 stainless steel .
Thermal Surface Assembly Dimensions: Diameter 50mm ± 2mm, Thickness: 6mm
Surface Texture: conforms to ISO 1302
2.3 Sprinkler System
The sprinkler system consists of a liquid supply device, two valves, and pipelines that guide the spray .
The direct volume of the valve is 1.0 cm³ ± 0.2 cm³.
Spray Pipe Dimensions: Length 250mm ± 5mm, outer diameter ≤ 4mm, inner diameter: 1.6mm ± 0.1mm
2.4 Temperature Control System:
A high-precision PID controller and thermocouples (usually welded to the back of the hot plate ) are used to achieve precise control and measurement of the hot surface temperature.
Temperature Setting Range: 0-1000℃
Accuracy: ±1℃
Resolution: 0.1°C
2.5 Number of Tests: 0-999 (can be set)
Test Interval Time: 0-999 seconds can be set
2.6. Power Supply: 220V 50/60HZ 2000W
2.7 Equipment Dimensions: 1200*800*1500mm
2.8 Standards Compliance: IEC 60335-2-40 Appendix KK, UL 60335-2-40