High and Low Temperature Humidity Test Chamber
Instrument Specifications
| Specification |
Details |
| Product Name |
High and Low Temperature Humidity Test Chamber |
| Instrument Model |
RSR1000-040PFE |
| Factory Number |
SHS20252072 |
| Production Date |
20250820 |
| Test Chamber Volume |
1000 L |
| Inner Box Size |
1000 mm × 1000 mm × 1000 mm (W×H×D) |
| Outer Box Size |
1275 mm × 2110 mm × 2100 mm (W×H×D) |
| Power Supply Voltage |
AC 480 V 60 Hz three phase five wire |
Standard Accessories
- One test chamber
- Power cord (extending 5 meters outside the device)
- Two silicone soft plugs (for test holes)
- Test specimen shelf with two layers
- Two sets of shelf hooks
- Special wet ball gauze (five strips)
- Operating instructions
- Certificate of conformity and warranty card
Main Technical Parameters
Temperature Parameters
| Parameter |
Value |
| Temperature Range |
-40℃ to 150℃ |
| Temperature Fluctuation |
≤ ±0.5℃ |
| Temperature Deviation |
≤ ±2.0℃ |
| Temperature Uniformity |
≤2.0℃ |
| Cooling Rate |
≥1.0℃/MIN |
| Heating Rate |
≥2.9℃/MIN |
Humidity Parameters
| Parameter |
Value |
| Humidity Range |
10% RH to 98% RH |
| Temperature Range of Controlled Humidity |
10℃ to 90℃ |
| Humidity Deviation |
≤ ±2.0%RH |
Note: The above parameters are measured under ambient temperature 25℃, relative humidity ≤85%RH, no heat source around, and no load in the test area.
Load Specifications
- Heat Source Load: No load
- Specimen Load: No load
Usage Environment Requirements
- Ambient temperature: 5℃-35℃
- Relative humidity: ≯85%RH
- Atmospheric pressure: 80kPa-106kPa
- Reserved space: ≮2 meters in front, ≮1 meter in the back and sides, ≮0.5 meters at the top
- Wastewater discharge: Drainage device or collection box within one meter from rear
- Keep away from direct sunlight, with smooth airflow, no strong vibration, heat sources, or high-concentration dust/flammable gases